MAINTENANCE
5-28 31 Dec 13
DE-7200VFDCentrifuge
ALARM AND FAULT MESSAGES
Cause Corrective Action
F2012 - Conveyor VFD Fault and Shutdown, Ref Troubleshooting Section of User Manual
Conveyor VFD fault Identify fault from operator control panel. Perform
indicated corrective action.
F2013 - Conveyor VFD Ground Warn Alarm, Ref Troubleshooting Section of User Manual
Conveyor VFD has detected high
current leakage to ground
Check condition of cable and all connections
between conveyor VFD and conveyor motor, and
correct any defects.
F2014 - Conveyor VFD In Phase Loss Alarm, Ref Troubleshooting Section of User
Manual
Conveyor VFD has detected unbalance
or partial loss of input three-phase
power
Check wiring between VFD and supply source;
correct defect(s).
Check supply power, and correct any defects.
F2015 - Conveyor VFD Power Loss Alarm, Ref Troubleshooting Section of User Manual
Incoming power loss or low voltage
supply
Check incoming power, and correct any defect(s).
F2016 - Conveyor VFD Undervoltage Alarm, Ref Troubleshooting Section of User Manual
Incoming power loss or low voltage
supply
Check incoming power, and correct any defect(s).
F2021 - Conveyor Motor Thermal Overload Check Conveyor Motor Temperature and
Thermistor
Conveyor drive motor shut down due to
high temperature or defective thermistor
in motor winding
Check motor temperature. Allow motor to cool, and
then check for continuity across thermistor leads. If
open circuit is found, replace motor.
Bad connection(s) Check connections at TB2-5 and 6, at
transmitter/intrinsic barrier, and at conveyor VFD.
Correct as required. Replace motor if no defects are
found.
F3001 - Pump Shutdown or Will Not Start Due to Dry Run Protection Alarm
Low tank level Check and re-fill feed tank sufficiently to close low
level switch.
F3002 - Flow Meter Flow Detection Error, Feed Pump has Shutdown
Insufficient pump output detected Check and re-fill feed tank sufficiently. Verify that
control valves (if any) are open to allow material to
flow.