INSTALLATION
3-10 01 Jan 14
DE-7200VFDCentrifuge
LOWER AND SECURE ROTATING ASSEMBLY
Note!
Do not discard the rubber strips between bearing pillow blocks, cover,
and base after removal. These strips must be re
-installed whenever the
machine is moved to prevent damage to the rotating asse
mbly bearings during
transit.
Following final positioning and leveling of the centrifuge, shipping blocks inserted between the
bearing pillow blocks and the base, as well as between the case and base, must be removed. The
shipping blocks prevent damage to the rotating assembly bearings during transit and should be
retained for future installation if the centrifuge is transported.
1. Remove all bolts securing top cover along both sides, but do not remove bolts on both ends of
cover.
2. Remove bolts securing bearing pillow blocks to base and bolts securing upper halves of pillow
blocks to lower halves (Figure 3-8).
Figure 3-8. Lowering and Securing Rotating Assembly